A practical job aid for vortex mixers and centrifuges in TB labs
A job aid turns a technical procedure into a short, usable set of actions that staff can follow at the bench. For tuberculosis laboratories, it should support safe specimen processing, reliable test results and consistent practice across permanent, rotating and newly trained personnel. The document must be easy to consult while wearing gloves and working inside a controlled laboratory area.
Vortex mixers and centrifuges are common preparation equipment, yet incorrect use can create aerosols, damage tubes, compromise sample integrity or produce an uneven sediment. A useful aid therefore needs to explain the purpose of each step, identify hazards, state when to stop, and direct staff to the laboratory’s approved standard operating procedure.
The GLI Quality Tool provides a suitable quality management context for this work. Its quality checklists can help a laboratory connect the job aid with equipment control, personnel competency, document management, safety, assessment and continual improvement.
Define the task and its limits
Start by stating exactly which TB specimens and procedures the job aid covers. For example, it may apply to the mixing and centrifugation of sputum or other respiratory specimens before a molecular, smear, culture or specimen-processing step. It should name the approved tube types, maximum volume, required rotor or bucket, validated speed and run time.
Avoid presenting the job aid as a substitute for the full procedure. Include a short scope statement such as: “Use this aid only with the validated TB specimen-processing method and current laboratory SOP.” Staff should know when a different instruction applies, such as for blood, liquid culture, decontamination, spill response or maintenance.
The document should also identify who may perform the task. Australian laboratories may have staff across metropolitan services in Melbourne, Sydney or Brisbane, as well as regional and remote facilities with different equipment models and staffing patterns. A clear scope reduces the risk that a worker transfers an instruction to an unvalidated centrifuge or a different specimen type.
Put safety before operation
The opening safety panel should tell staff to check the risk assessment, wear the required personal protective equipment and work in the designated containment area. Processing potentially infectious TB material should follow the laboratory’s biosafety requirements, including use of a biological safety cabinet where specified. Gloves, protective clothing and eye or face protection should be selected according to the activity and local risk assessment.
For the vortex mixer, the aid should say to keep the tube capped, inspect it for cracks or leakage, hold it securely and use the validated mixing setting. Staff should never vortex an open tube or press a damaged container against the mixer. If a tube leaks, breaks or becomes visibly contaminated, they should stop and follow the spill and exposure procedure rather than attempting to continue.
For centrifugation, describe the use of sealed safety cups or aerosol-tight rotors when required by the risk assessment. Buckets must be balanced by mass, not simply by visual appearance, and placed opposite one another according to the manufacturer’s instructions. The lid must be locked before the run begins. If unusual vibration, noise or movement occurs, staff should stop the instrument if safe to do so and report the fault.
Make the vortex steps unambiguous
A practical job aid should show the sequence in the same order used at the bench. Staff first confirm the specimen identifier against the request or worksheet, check the tube condition and make sure the cap is secure. They then place the tube on the mixer according to the approved method, using the validated time and speed rather than an informal phrase such as “mix well.”
The instruction should explain whether the tube is mixed continuously or in short pulses, and whether the operator must keep a finger on the cap. Where the specimen is viscous, the aid can state what adequate mixing looks like without requiring the operator to open the tube. It may describe a uniform suspension or the absence of visible settled material, provided that this criterion has been approved by the laboratory.
Timing matters. A wall-mounted timer, instrument timer or validated electronic system is preferable to estimating duration. The job aid should direct staff to return the tube to its correct rack position immediately after mixing, maintain the identifier’s visibility and proceed without unnecessary delay if the test method requires prompt centrifugation.
Set out centrifuge loading and unloading
The centrifuge section should begin with a pre-use check. Staff inspect the chamber, rotor, buckets, seals and adapters for contamination, cracks, corrosion or damage. They confirm that the selected rotor is approved for the instrument, that the lid and safety lock function correctly, and that the programmed speed and time match the current SOP.
Explain balancing with a specific example rather than relying on memory. Two tubes of different volumes may not be balanced by appearance; the laboratory should specify how to use a balance or matched tubes and how to prepare a suitable balance tube. Tubes must be seated fully in their adapters, and unused positions should be managed according to the centrifuge manufacturer’s instructions.
The job aid should tell staff to close and lock the rotor or safety cups before placing the assembly in the chamber. After the cycle, they should wait for the rotor to stop completely before opening the lid. If an aerosol-tight system is used, opening safety cups should occur in the appropriate containment device or area and in accordance with the risk assessment. This is especially important when a service receives specimens by courier from remote Queensland or Western Australia and a tube may have shifted or leaked in transit.
Protect sample integrity and traceability
Centrifugation parameters affect the distribution of organisms, cells, debris and liquid within a specimen. The job aid should therefore connect each specimen type with its validated relative centrifugal force or revolutions per minute, duration, temperature where relevant and post-spin handling instruction. If the equipment displays RPM but the SOP specifies RCF, provide the approved conversion or label the correct programme on the instrument.
Staff need clear directions about the pellet and supernatant. State which fraction is required, how to remove it without disturbing the pellet, and where it must be transferred. The instruction should prohibit vigorous movement after centrifugation when a settled fraction must remain intact. Any unacceptable outcome, such as a broken tube, visible leakage, incorrect programme or disturbed pellet, should trigger documented assessment by an authorised person.
Identification and chain of custody are part of sample quality. Record the specimen or accession number, operator, date, time, instrument identity and any deviation required by the laboratory system. When TB results may contribute to legal proceedings, laboratories should follow a documented chain of custody guide so that movement, access and processing records remain defensible.
Add cleaning, faults and emergency actions
A good aid explains what the operator does after every run. Staff should remove racks and tubes safely, inspect the chamber for contamination and clean the vortex mixer, rotor, buckets and work surface with the laboratory-approved disinfectant. The contact time, compatibility limits and disposal method must come from the relevant SOP or manufacturer’s guidance; bleach or another chemical should not be used automatically on equipment that it may damage.
The fault panel should be brief and visible. It can cover imbalance alarms, failure to start, unusual vibration, broken tubes, leakage, power interruption and a lid that will not open. Staff should not bypass an interlock, force a lid or restart equipment after a suspected mechanical failure. The aid should state whom to notify, how to label and isolate the instrument, and where to record the event.
In Australia, this wording should align with the organisation’s work health and safety system and applicable state or territory requirements, including duties under the model Work Health and Safety framework where adopted. A regional service may have a longer wait for a biomedical technician than a central laboratory in Adelaide or Perth, so the escalation pathway should include local contacts, equipment isolation and safe specimen storage.
Validate, train and review the aid
Before release, compare every parameter with the current manufacturer manual, risk assessment, validation record and approved SOP. A laboratory should test the draft with the actual vortex mixer and centrifuge, including the safety cups, adapters, controls and display language in use. Photographs or simple diagrams can help staff recognise the correct rotor, tube position and balance arrangement, but images must match the local equipment.
The supervisor should observe each user performing the task and document competency, including safe loading, correct settings, traceability and response to an abnormal event. A short read-and-demonstrate assessment is more useful than a signature alone. Refresher training should occur after a significant incident, equipment replacement, procedural change or observed error.
Keep the controlled version near the point of use, with a document number, approval date, review date and owner. Remove superseded copies from benches and shared drives. The GLI guidance on planning a safer lab layout can support decisions about separating clean and dirty workflows, positioning equipment and reducing cross-contamination risks.
A job aid becomes valuable when staff can use it quickly without guessing. Write short action statements, define technical terms such as RCF, use the same names as the instrument controls, and include stop points for unsafe conditions. Invite feedback from bench staff, document errors and near misses, and revise the aid through the laboratory’s document-control process.
Download the relevant GLI Quality Tool materials, adapt the steps to your validated TB workflow and test the finished aid at the bench before approval. A clear local instruction can strengthen staff competency, reduce preventable exposure and support dependable results across Australian laboratory services.