Writing a TB lab standard method for annual calibration checks

Writing a TB laboratory standard method for verifying centrifuge speed and timer calibration annually should turn a routine check into defensible laboratory evidence. Identify the instrument and what is being checked, specify a traceable reference, and state what happens when readings fail. Distinguish an internal verification from calibration performed by a qualified calibration service.

Start with the equipment list rather than a generic instruction. Name each centrifuge, rotor, timer and reference device, including asset number, location and normal operating range. State whether speed is controlled in revolutions per minute (RPM) or relative centrifugal force (RCF), and which routine processing times are critical. This prevents a check of the display alone being mistaken for verification of actual rotor performance.

The procedure should sit inside the laboratory quality management system (QMS), with document control, training records, equipment maintenance and nonconforming work processes linked. The GLI Quality Tool’s four-phase roadmap and twelve Quality Systems Essentials provide a useful framework: equipment, documentation, assessment and continual improvement all shape how the annual evidence is managed.

Australian laboratories range from metropolitan public-health units in Melbourne, Sydney and Perth to regional and remote services around Darwin and Alice Springs. Staffing, engineer access and budgets differ sharply. Where accreditation applies, NATA assessment commonly uses ISO/IEC 17025 requirements for equipment, traceability and nonconforming work. Many teams buy calibration services or replacement parts through national suppliers, so maintenance and ownership of reference equipment must be clear. Staff may say they will “sign off” a check only when the evidence is complete and traceable.

Set the scope and acceptance criteria

Name the SOP accurately. “Annual verification of centrifuge speed and timer accuracy” is usually more precise if staff compare the instruments with traceable references and apply predetermined limits without adjusting them. Reserve “calibration” for a process that establishes measurement traceability, including adjustment where required. If an external calibration certificate is supplied, identify it and explain whether it supplements or replaces the internal annual check.

Set acceptance criteria before testing. Base speed and timing tolerances on the manufacturer’s instructions, the laboratory’s validated method, risk to results and relevant accreditation requirements; do not invent a universal percentage. Express RCF consistently, noting that it depends on rotor radius. For timing, state the permitted error at each routine interval and whether equipment may remain in service when the error is within tolerance. Include the reference device’s identification, calibration status and expanded uncertainty where relevant.

The GLI Quality Tool’s laboratory quality checklists provide a practical starting point for locating the equipment, documentation and assessment requirements that affect this task. Adapt them to the laboratory’s own risk profile rather than copying them unchanged, then record the selected limits in the controlled procedure.

Build the method for safe, repeatable testing

A controlled SOP should have a unique number, version, owner, approver, effective date and review date. Include purpose, scope, definitions, responsibilities, references, equipment, acceptance criteria, step-by-step instructions, records and actions for failed checks. Reference the centrifuge and timer manuals, maintenance history and risk assessment. Write each step as an observable action, including what the operator sees, measures, records and signs.

Because TB material may be present, the method must pair equipment verification with laboratory biosafety controls. Inspect the rotor, lid, balance, tube ratings and containment before operation; use compatible, correctly balanced loads and follow the approved decontamination and spill procedures. Training should make clear that a centrifuge is not a biological safety cabinet and that unexpected noise, vibration, heating or a damaged rotor requires the equipment to be stopped, tagged and assessed.

Name the operator and independent reviewer. List PPE, the approved reference tachometer or stopwatch, timer, rotor-radius information and forms. For regional and remote settings, include a pre-use visual check, authorised remote review and arrangements for sending equipment for calibration. State who bears the cost and downtime so the schedule remains achievable.

Perform the annual checks

Before testing, confirm the centrifuge and timer identifiers, reference-device calibration certificates, environmental conditions, rotor condition and latest maintenance status. Work on a clean, level bench with uninterrupted time for the test. Use a manufacturer-approved load or the load defined by the laboratory’s method. Record actual settings rather than assuming the instrument was operating as displayed.

For centrifuge speed, run at the relevant operating speed and compare an independent tachometer with the set RPM. If the method uses RCF, calculate it from measured RPM and rotor radius: RCF = 1.118 × 10⁻⁵ × r (cm) × RPM². Test routinely used speeds as appropriate, allowing the rotor and load to stabilise. For timer accuracy, compare the interval with a traceable reference timer over selected durations, repeating short and long routine settings and recording readings.

Check Centrifuge speed verification Timer accuracy verification
Objective Confirm the rotor runs at the specified RPM or calculated RCF Confirm the displayed interval matches elapsed time
Reference Calibrated optical or laser tachometer with valid traceability Traceable stopwatch or reference timer with valid calibration
Operating condition Approved rotor, balanced compatible load and stable operating speed Instrument set to a defined duration under normal use conditions
Evidence Instrument and rotor identifiers, set RPM, measured RPM, calculated RCF and acceptance limit Tested settings, reference elapsed times, displayed intervals, differences and acceptance limit
Decision Release when each result meets the stated criterion Release when each timing difference meets the stated criterion

After either check, calculate the difference from the reference, compare it with the predefined criterion and record the result without overwriting raw readings. A failed check does not become acceptable because the instrument later appears normal. Place the affected equipment under the laboratory’s nonconforming-work control until investigation, adjustment, retesting and formal release are documented.

Keep records and manage failures

Retain the completed worksheet, reference-device certificates, maintenance history and reviewer sign-off. Records should show the date, operator, equipment location, settings, reference-device identification, actual readings, calculated differences, decision and linked corrective-action number. Store them under document control so an auditor can trace each result to the personnel, method version and equipment used.

When a result is outside tolerance, stop using the equipment for the affected application and assess whether any test results generated since the last acceptable check require review. Define immediate correction, root-cause analysis and corrective action proportionately. Typical causes include a loose or failed sensor, timer setting error, rotor mismatch, software configuration issue or worn equipment. Verify effectiveness through retesting; do not use retraining as a substitute for repair when equipment is defective.

A concise record is easier for staff to follow and for an assessor to audit. The GLI Quality Tool feedback page can receive comments about unclear instructions, missing fields or checks that do not fit the laboratory workflow. Review recurring deviations during quality meetings and feed confirmed process changes into document control, so improvements affect routine practice rather than remaining discussion points.

Review, approve and put the method into practice

Treat the annual verification as a required control, not a substitute for preventive maintenance or manufacturer-specified calibration. Review the frequency against equipment use, previous drift, transport, repairs and the impact on TB test results. Even when policy requires annual checks, a heavily used centrifuge or one moved between sites may need interim verification. Record the rationale for any additional checks and the chosen schedule.

Set a formal review date and also trigger an out-of-cycle review after a failure, repair, software update, rotor change or change in critical processing time. Analyse trends in RPM deviation, timing error and corrective action across the laboratory. A rising drift that remains within tolerance still warrants investigation; the aim is continual improvement and early intervention rather than repeated annual surprises. Use the findings to refine maintenance, training and preventive controls.

Confirm that staff can perform the check competently and that the current SOP matches what they actually do. Competency assessment may include a practical demonstration, observation and a review of the first completed worksheet. The authorised approver should confirm that acceptance criteria remain suitable, references are current, forms match the procedure and all deviations are closed. Then issue the approved version, withdraw obsolete copies and announce what changed.

Adopt the approach as a controlled, risk-based SOP with two linked records: the procedure and the completed annual verification form. Before implementation, enter the equipment and reference-device identifiers, define the speed and timing tolerances, assign operator and reviewer roles, link biosafety requirements and specify the response to failure. Use the GLI Quality Tool checklists and phase-specific guidance to confirm that equipment, documentation, safety, assessment and improvement activities are in place. Complete the annual check, record the evidence, resolve any nonconformance and release the equipment for use only when the result is defensible.