GLI GLI Quality Tool
GLI Quality Tool — Version 2.0

A practical guide to using biological safety cabinets in TB laboratories

A biological safety cabinet (BSC) is a primary containment device that protects laboratory staff, the surrounding environment, and, in some configurations, the work from infectious aerosols. For tuberculosis laboratories, correct cabinet use is essential during procedures that may generate aerosols, including specimen processing, vortexing, aliquoting, centrifuge tube handling, and preparation of cultures or molecular test materials.

A BSC cannot compensate for poor technique, inadequate training, or an unsuitable room. It works as part of a wider biosafety system that includes risk assessment, respiratory protection policies, equipment maintenance, waste management, incident reporting, and documented procedures. Staff should therefore understand both the cabinet’s airflow and the laboratory practices that preserve containment.

This guide is intended to support routine work in laboratories with different levels of infrastructure. It can be adapted into training material or a standard operating procedure (SOP), while local regulations, manufacturer instructions, and the laboratory’s risk assessment remain authoritative.

Select and prepare the right cabinet

Most TB diagnostic procedures involving potentially infectious patient material should be performed in a certified Class II BSC. This type of cabinet draws room air through the front opening, directs airflow across the work area, and passes exhaust air through high-efficiency particulate air (HEPA) filtration. The exact protection provided depends on the cabinet class, configuration, installation, and certification status.

A clean bench or laminar airflow cabinet is not a substitute for a BSC. Clean benches protect the product by directing air toward the user, but they do not protect staff from infectious aerosols. Chemical fume hoods are also designed for a different purpose. Before beginning work, confirm that the cabinet is intended for microbiological containment and that its certification label is current.

Position the cabinet away from doors, windows, fans, air-conditioning outlets, and busy walkways. Moving air can disrupt the protective inflow at the front opening. The room should provide enough space for safe access, cleaning, and servicing. Do not use a cabinet that has an alarm, visible damage, blocked grilles, failed airflow, or an expired certification date unless an authorized engineer has assessed it.

Start every session with a deliberate check

Only trained and authorized personnel should operate a BSC. Review the relevant SOP, confirm the procedure and materials needed, and remove unnecessary items from the work area. Keep the front and rear grilles clear; blocked grilles can interrupt airflow and create turbulence. Place materials so that clean items enter before contaminated items, and avoid reaching repeatedly across infectious material.

Switch on the cabinet according to the manufacturer’s instructions and allow the recommended purge time before work begins. Check the airflow indicator and listen for unusual sounds. If the cabinet has an alarm or the airflow reading is outside the acceptable range, stop and report the problem. Never silence an alarm simply to continue testing.

Wear the personal protective equipment required by the risk assessment. A laboratory coat or gown, gloves, and eye or face protection may be necessary, while respiratory protection may be required for specific procedures or room conditions. Gloves should be changed when damaged or visibly contaminated, and gloved hands should never touch phones, door handles, notebooks, or personal items.

Work smoothly inside the airflow zone

Arrange the cabinet from clean to contaminated: supplies and sterile materials on one side, the active work in the center, and waste containers on the opposite side. Keep frequently used items within reach, because unnecessary arm movements can disturb airflow. Work several centimeters inside the cabinet rather than directly at the front grille, where inflowing air is most vulnerable to disruption.

Move arms slowly and steadily, entering and leaving the cabinet in a straight line. Do not use open flames, because flames generate turbulence and can damage filters or cabinet components. Use an approved alternative when sterilization is needed. Keep containers closed whenever possible, and cap tubes before removing them from the work zone.

Centrifugation requires particular care. Use sealed safety cups or rotors, load and unload them inside the BSC when appropriate, and allow aerosols to settle according to the local SOP before opening. If a tube breaks or a spill occurs, keep the cabinet running, pause activity, and follow the documented spill response. A useful SOP should specify disinfectant, contact time, exposure precautions, cleanup materials, and incident reporting.

Activity Safe practice inside the BSC Frequent error to prevent
Specimen opening Open containers slowly and keep them away from the front grille Working at the cabinet edge
Pipetting Use low-aerosol technique and keep tips below the breathing zone Forceful expulsion or repeated rapid movements
Vortexing Use sealed tubes and allow aerosols to settle Opening tubes immediately after mixing
Centrifuge handling Use sealed cups or rotors and inspect them for damage Opening an unsealed rotor outside containment
Waste collection Use a stable, closable container within reach Overfilling or reaching across contaminated items
Spill response Keep airflow on and follow the spill SOP Switching off the cabinet during cleanup

Clean, disinfect, and manage waste correctly

At the end of work, close or secure specimen containers, remove waste using the established workflow, and disinfect the interior surfaces. Begin with the least contaminated areas and work toward the most contaminated area, using a disinfectant that is effective against the organisms involved and compatible with the cabinet surface. Observe the full contact time; wiping a disinfectant away immediately may reduce its effectiveness.

Do not place large equipment, absorbent materials, cardboard, or unnecessary supplies in the cabinet. These items can obstruct airflow and make effective cleaning difficult. Ultraviolet lamps, where present, should not be treated as a replacement for surface disinfection. They have limitations, require maintenance, and can expose staff to harmful radiation if used incorrectly.

Waste segregation should begin at the point of generation. Sharps, liquid waste, contaminated plastics, and reusable materials need separate handling according to local policy. The GLI resource on waste sorting procedures can help staff connect cabinet practices with a clear, documented waste pathway. Never overfill containers, and close them before removal from the cabinet.

After cleaning, inspect the work zone, remove gloves safely, wash hands, and record any spill, equipment fault, or deviation. If contamination may have reached a grille, filter area, or external surface, do not improvise a repair. Restrict use and notify the designated biosafety or maintenance lead.

Verify performance and respond to problems

A BSC must be certified after installation, relocation, repair, filter replacement, or other events that could affect performance, and at scheduled intervals defined by national requirements or the manufacturer. Certification normally includes airflow measurements, HEPA filter integrity testing, smoke visualization, and checks of alarms and cabinet function. Local staff should know where the certificate is stored and when recertification is due.

Daily or session-based checks are also important. A simple log can include date, operator, cabinet identification, airflow status, alarm status, cleaning completed, and any abnormal observation. These records provide evidence that the cabinet was used correctly and help identify recurring faults, training gaps, or maintenance delays.

Quality indicators can make this information easier to act on. For example, a laboratory may track overdue certification, failed pre-use checks, spill events, cleaning-log completion, or staff competency assessments. A quality indicator dashboard can turn these observations into a monthly review rather than leaving them scattered across paper records.

If an airflow alarm activates during work, stop manipulating infectious material, secure containers if it is safe to do so, and inform the supervisor. Do not continue because the procedure is almost finished. If a spill occurs, keep people away from the area, allow aerosols to settle when required by the SOP, and document the event after cleanup and medical evaluation where indicated.

Build competence through observation and review

Reading an SOP is not enough to demonstrate competence. New users should receive an explanation of cabinet principles, observe an experienced operator, practice under supervision, and complete a documented assessment. Competency checks should cover setup, material placement, pipetting, centrifuge handling, spill response, decontamination, waste removal, and actions during an alarm.

Refresher training is needed after a serious incident, a major procedural change, a long absence from cabinet work, or evidence of repeated errors. Supervisors can use direct observation, checklists, and short practical exercises. Training should also explain why behaviors matter, so staff understand that rapid movements, blocked grilles, and clutter can weaken containment even when the cabinet appears to be functioning.

Proficiency testing and internal quality assessment can reveal whether safe cabinet practices support reliable TB testing. When a result or process indicates a problem, laboratories should investigate causes rather than focusing only on individual blame. The guidance on reviewing proficiency results offers a useful model for documenting findings, corrective actions, and follow-up verification.

Make safe cabinet use part of quality management

The laboratory’s quality management system should connect biosafety cabinet operation with document control, personnel records, equipment management, assessment, and continual improvement. The cabinet SOP should identify responsibilities, required PPE, startup and shutdown steps, work-zone layout, cleaning agents, waste routes, emergency actions, maintenance contacts, and records to retain.

Keep the SOP accessible at the workstation, but do not rely on a document alone. Post a simple airflow and work-organization diagram if it helps users arrange materials consistently. Supervisors should review logs regularly, discuss deviations during team meetings, and ensure corrective actions have an owner and a completion date.

Practical habits that support reliable containment include:

A well-used BSC is a critical barrier in TB laboratory work, but its protection depends on consistent human behavior and a functioning management system. Lab staff should apply these practices during every session, supervisors should observe and reinforce them, and quality teams should use records and indicators to improve the process over time. Download or adapt the relevant GLI Quality Tool materials, incorporate local requirements into the SOP, and begin routine competency checks at the next team training session.