Designing a TB lab proficiency panel for rapid molecular testing
A TB laboratory proficiency panel examines whether a rapid molecular testing service can produce accurate, reliable and clinically useful results across the full pathway: specimen collection, transport, extraction, amplification, control acceptance, interpretation and reporting. Its value lies in revealing system weaknesses that an instrument’s internal controls may not expose, particularly in laboratories handling mixed workloads and variable staffing.
The GLI Quality Tool provides a four-phase roadmap organised around 12 Quality Systems Essentials, with practical support for safety, personnel, equipment, documentation, assessment and continual improvement. A proficiency panel should therefore function as a quality-management activity, not an isolated test event. This guidance addresses designing a TB lab proficiency panel for rapid molecular testing and interpretation of indeterminate results.
Define the purpose and scope
Start with a precise statement of purpose and scope. Name the assay platforms, such as Xpert MTB/RIF or Xpert MTB/RIF Ultra, the target analytes, specimen types, operator groups and reportable outcomes. If the panel assesses detection of Mycobacterium tuberculosis complex and rifampicin resistance, say so; do not imply that it substitutes for culture confirmation, antimicrobial-susceptibility testing or assessment of the whole clinical pathway.
Build a balanced challenge set rather than a collection of easy positives. Include negative specimens, clear positives, low-burden samples near the relevant limit of detection, rifampicin-resistant challenges, low-volume or inhibited specimens and validated control-failure challenges. Add non-tuberculous mycobacteria or cross-reactivity challenges only when suitable verified material is available. Size the panel against assay volume and risk. For every specimen, record the expected result, acceptable range, reference method and device-specific decision rule so scoring reflects agreed evidence. Where an external quality assessment programme already supplies material, apply its rules without alteration.
Select and verify specimen materials
Use noninfectious manufactured controls or verified, thoroughly inactivated material where possible to limit biosafety risk. Confirm that processing and transport do not alter the expected outcome. Include an extraction control and, if the objective requires it, a validated inhibitor challenge; poor handling should not be used as an artificial substitute for a genuine control failure. Align specimen type, container, volume, storage and transport conditions with routine diagnostic samples so the exercise tests operational readiness.
Australian distribution requires careful planning between Melbourne, Sydney, Brisbane and regional or remote services. Heat, flight schedules and road delays can affect samples, so define temperature and time limits, tracked courier procedures and rules for quarantining material with an excursion. International procurement also requires checking Australian Border Force and state or territory biosecurity requirements before any infectious or regulated shipment. Inactivated material may follow different rules, so obtain written import conditions rather than assuming customs clearance.
Control testing conditions and performance
Blind the panel with neutral codes, conceal concentrations and randomise processing order; keep the expected-result key under separate control. When comparing operators or shifts, rotate equivalent challenges rather than repeat one predictable specimen. Testing staff should use the normal extraction pathway, internal quality control, maintenance schedule and reporting interface. Before release, confirm calibration, preventive maintenance, power stability, environmental conditions and secure data storage; a contrived workflow should not be mistaken for routine performance.
Record specimen receipt, integrity, instrument, reagent lot, operator, controls, result, repeat and report. Keep analytical failure separate from an operator mistake: an invalid control caused by a verified reagent fault should not be scored as poor technique. Conversely, correctly detecting a target while misclassifying an invalid result, repeating it improperly or failing to communicate it represents a real competency gap.
Interpret indeterminate results
Define indeterminate before issuing the panel because laboratories may use it for an invalid assay, instrument error, borderline signal, insufficient material or unresolved clinical concern. Follow each manufacturer’s instructions for use and distinguish a reportable molecular finding from a run failure. Where a valid target result can coexist with a failed control, apply the platform’s exact repeat and reporting rule rather than improvising an interpretation.
Set out permitted repeats, required fresh extraction or cartridge, escalation to a senior scientist, and when recollection or a different assay is needed. The expected response should include control review, investigation of pre-analytical factors, repeat testing, documentation and an appropriate preliminary or final report. Score both the result and the quality of the response. During routine patient care, notify the treating clinician promptly and follow state or territory tuberculosis notification requirements through the relevant communicable disease team; proficiency-panel scoring must never delay this action.
Score findings and document action
Score the expected result, observed result, reportability, repeat, deviation, root cause and corrective action. Distinguish analytical, interpretive, reporting and administrative errors. Use sentinel rules so a missed low-burden target, false rifampicin-resistance call or mishandled invalid result triggers case review regardless of the overall percentage. Assign an owner and due date to every action, then review instrument logs, reagent records, training, environmental data and specimen handling before deciding whether the cause lies with material, equipment, procedure, competence, communication or panel design.
Use the GLI Quality Tool to connect the exercise with the laboratory quality management system. Staff can use phase-specific checklists to verify documentation, equipment, staffing, safety, assessment and improvement controls. Select checklists relevant to the current phase and retain evidence that identified gaps have been assigned, completed and reviewed. A failed challenge should be repeated only after corrective action has been tested and its effectiveness recorded.
Build continual improvement into each round
Repeat proficiency panels at an interval matched to assay volume, clinical risk and operational change. Rotate specimen types, operators and failure modes so staff cannot learn a fixed answer key. Track recurring invalid rates, turnaround time, repeat attempts, specimen rejections and the interval between an unresolved result and clinical notification. Use de-identified cases in routine quality meetings to reinforce decisions and revise standard operating procedures, rather than presenting sites as a punitive league table.
Australian laboratory networks may combine large metropolitan services with regional hubs and near-patient testing, including services supporting Aboriginal and Torres Strait Islander communities. Interpret differences in transport, staffing, cultural safety and technical support before comparing results, and involve local clinical leaders where access or communication affects testing and notification. Audit routine reports and selected challenge repeats to confirm that improvements persist. Procurement through state tenders, hospital purchasing or the Australian commercial pathology market should also keep panel material aligned with the assays and service contracts actually in use.
Start by agreeing on the panel’s purpose, budget and responsible technical lead, then obtain or prepare verified material through a laboratory with appropriate quality assurance. Develop the expected-result key in advance, pilot the instructions with one operator outside the scored group, and have the clinical and quality teams review the escalation pathway. Keep the exercise non-punitive and document its approval before distribution.
Teams ready to begin can download quality materials to access the relevant templates and instructions, then map the exercise to the GLI Quality Tool’s phase and Quality Systems Essentials. Assign owners for scoring, corrective action and the next panel date, and protect time for staff to interpret the results and apply the lessons to routine practice.