GLI GLI Quality Tool
GLI Quality Tool — Version 2.0

Building Reliable Internal Quality Control for Xpert MTB/RIF

A dependable internal quality control system helps a tuberculosis laboratory detect problems with Xpert MTB/RIF testing before inaccurate results affect patient management or surveillance data. The system should cover the complete testing process, from specimen receipt and cartridge storage to result verification, reporting, and corrective action.

Xpert MTB/RIF and Xpert MTB/RIF Ultra are cartridge-based molecular assays with built-in controls, but those controls do not replace a laboratory-managed quality system. Internal quality control should also monitor operator practice, environmental conditions, equipment performance, specimen quality, documentation, and recurring error patterns.

The most effective approach is practical and risk-based. Laboratories should define what must be checked, assign responsibility, establish review intervals, and retain enough evidence to show that testing remains controlled over time. The four-phase quality improvement model and its Quality Systems Essentials can help organize this work in laboratories with different staffing levels and resources.

Define the purpose and scope

Begin with a written procedure describing the purpose of internal quality control for Xpert testing. State which instruments, assay versions, specimen types, operators, testing sites, and reporting processes are covered. The procedure should also define related terms, including invalid, error, no result, instrument failure, contamination concern, and rejected specimen.

Quality control objectives may include confirming that cartridges and reagents are stored correctly, ensuring operators follow the approved workflow, identifying unusual result distributions, monitoring instrument error rates, and verifying that corrective actions are effective. These objectives should be linked to risks identified by the laboratory rather than copied into a generic checklist.

The user instructions for the GLI Quality Tool can help staff navigate phase-specific materials and apply the quality systems approach consistently. The laboratory can then adapt the guidance to local policies, national TB algorithms, manufacturer instructions, and available resources.

Set control materials and frequencies

Xpert systems include internal process controls within cartridges, such as checks related to sample processing and amplification. These controls are important for validating each individual test, but laboratories should also establish oversight controls that examine the broader testing system. A positive control can help demonstrate that the assay detects target material, while a negative control can help identify contamination or problems in preparation and handling.

The use of additional control material must follow the current manufacturer instructions and national guidance. Laboratories should never modify cartridge handling or introduce materials into a cartridge unless the approved procedure specifically allows it. Where external positive or negative materials are used, document their source, concentration or expected reactivity, storage conditions, expiration date, and acceptance criteria.

Control frequency should reflect risk, workload, and the laboratory’s experience. A new instrument, new lot, major maintenance event, relocation, software update, or prolonged interruption may require enhanced checks. Routine monitoring can include daily review of instrument status and results, periodic testing of control materials, and scheduled analysis of error and invalid rates. Downloadable quality resources can support the creation of forms, checklists, and controlled records.

Standardize the daily testing workflow

A robust internal quality control system begins before the cartridge is inserted. Staff should verify specimen identification, request details, sample suitability, cartridge lot and expiry, storage conditions, instrument availability, and biosafety precautions. The person performing the check should record exceptions rather than relying on memory or informal communication.

At the instrument, operators should follow the approved steps for sample transfer, cartridge loading, barcode entry, test initiation, and result review. Work areas should be arranged to reduce spills, mix-ups, and contamination. Instruments need routine cleaning and maintenance according to manufacturer requirements, with service events and unresolved faults recorded.

Quality checkpoint Routine evidence Warning sign Required response
Specimen identification Request form, barcode, accession record Mismatch or missing identifier Stop testing and resolve identity before proceeding
Cartridge control Lot, expiry, storage record Damaged, expired, or poorly stored cartridge Quarantine and investigate the affected stock
Instrument status Run log, maintenance record Repeated errors or failed self-check Remove from service when necessary and escalate
Test result Electronic or paper result record Invalid, error, no result, or unexpected pattern Follow repeat and investigation procedure
Environmental conditions Temperature or site monitoring record Conditions outside the defined range Assess impact and document corrective action
Operator practice Observation or competency record Deviations from the approved workflow Provide feedback, retraining, or reassessment

The table should be converted into a usable local form rather than treated as a substitute for a procedure. Records need enough detail to connect a result with the specimen, operator, instrument, cartridge lot, date, and any action taken.

Interpret results and investigate exceptions

A valid positive or negative result should be reviewed within the laboratory’s reporting process, including any required clinical or algorithm-based interpretation. An invalid, error, or no-result outcome should trigger a defined response. The response may include checking specimen volume and quality, reviewing instrument messages, assessing cartridge integrity, repeating the test when permitted, and contacting a supervisor.

Repeat testing should not be used automatically to make an inconvenient result disappear. The reason for the repeat, the original outcome, the repeat outcome, and the final disposition should all be documented. If a repeat produces a different result, the discrepancy requires investigation rather than simple replacement of the first record.

Trend analysis adds value beyond individual run review. Calculate error, invalid, and no-result proportions by instrument, operator, cartridge lot, specimen type, and time period when the data volume supports it. A sudden increase may indicate instrument wear, environmental stress, poor specimen collection, storage problems, operator drift, or a faulty lot. Define alert thresholds locally and review them at a regular quality meeting.

Connect records, competency, and corrective action

Internal quality control depends on traceable records. A useful record set includes the standard operating procedure, daily equipment checks, temperature logs, cartridge inventory, control results, run reports, maintenance records, nonconformity forms, corrective actions, and staff competency assessments. Electronic records should be protected from unauthorized changes and backed up according to laboratory policy.

Competency assessment should cover the complete Xpert workflow, not just the ability to operate the instrument. Staff should demonstrate specimen acceptance, biosafety, cartridge handling, result interpretation, documentation, troubleshooting, and referral or escalation procedures. Assessment is needed at induction and periodically thereafter, as well as after major procedural changes or evidence of repeated errors.

When a nonconformity occurs, use a structured investigation. Identify what happened, when it began, which results or specimens may be affected, and whether a patient or surveillance report requires review. Corrective action may involve retraining, instrument service, stock quarantine, workflow redesign, environmental control, or revision of a form. Effectiveness must be checked later; closing an action because training was delivered does not prove that the problem has been resolved.

Use a phased quality improvement approach

Laboratories do not need to implement every quality activity at once. A phased approach can start with essential controls: an approved Xpert procedure, clear responsibilities, basic equipment and temperature records, result review, and a process for investigating invalid and error outcomes. Once these elements are stable, the laboratory can strengthen trend analysis, competency systems, internal audits, and management review.

The GLI Quality Tool organizes improvement around twelve Quality Systems Essentials, including personnel, equipment, documents and records, assessment, process management, and continual improvement. Its downloadable materials can help a laboratory map existing practices, identify gaps, and assign actions without losing sight of daily testing demands.

Management review should examine whether the internal quality control system is producing useful information. Review the number and type of exceptions, unresolved actions, instrument downtime, stock problems, staff competency findings, and relevant external quality assessment results. Use these findings to set measurable priorities for the next review period.

Practical recommendations for implementation

A laboratory can turn the principles above into a manageable routine by assigning ownership and making each control visible in the workflow.

Start with a short baseline review of current Xpert records, storage conditions, instrument logs, and staff practices. Then select a small number of corrective actions, assign deadlines, and measure whether error patterns and documentation gaps improve. This creates a quality control system that supports safe, consistent TB diagnosis while remaining realistic for the laboratory’s resources.

Use the GLI resources to structure the assessment, authorize the revised procedure, train the team, and begin routine review of evidence. A well-maintained internal quality control system turns each Xpert run into part of a larger, traceable process for protecting patients and strengthening tuberculosis laboratory services.